Simulator MS400 Patient Multiparameter

R21 469,90

Add R2 000,00 to cart and get free shipping!

Simulator MS400 Patient Multiparameter

MS400 Multi-Parameter Patient Simulator – ECG, Arrhythmia, IBP, Respiration, Temprature

MS400 Multiparameter Simulator is a portable device, which can provide comprehensive physiological waveform to content general test and train requirement. It can check the monitor rapidly near the bed, also can test the quality warranty of ECG equipment, monitor and remote therapy equipment, At the same time, it can be used as the training of distinguishing from arrhythmia waveforms to the nurses.

â– 3.5-inch color TFT touch-screen, 320*240 pixels, and friendly operation interface.
â– Display the output waveforms synchronously, which is convenient for user to
compare with the test equipment.
â– It integrates the functions of SD socket interface; user can output the
custom waveform by this device.
â– Built-in templates list to provide the waveform that is expressed by each
preset sequence   number, which is convenient for searching.
â– Two operation interfaces: Chinese and English.
â– Operate with keystroke and touch-screen expediently and rapidly.
â– Built-in 3.7V/4000mAh (Rechargeable alkaline battery).
â– Optical external power: AC 220V, 50Hz or AC 120V, 60Hz

SKU: SIM400MP Categories: , Tags: ,
Description

Simulator MS400 Patient Multiparameter

ECG/Arrhythmia
ECG
â– Conventional: All leads have the same voltage with the II-lead as the reference
â– Amplitude: 0.5,1.0,1.5 and 2.0mV
■The accuracy of Amplitude: ±2%
â– ECG Rate:
15,20,25,30,40,60,80,100,120,140,160,180,200,220,240,260,280,300,320,340      and 350 BPM
â– ECG artifacts: 50 or 60 Hz, muscle, baseline drift and breathing
â– The elevation of the ST segment: -0.8 mV to +0.8 mV, the step is 0.1 mV
â– The default: 80 BPM,1mV
Arrhythmia
â– Atrial: Atrial Fib(coarse), Atrial fib(fine), Atrial Flutter, Sinus Arrhythmia,
Missed Beat, Atrial
â– Tachycardia, Nodal Rhythm, Supraventricular Tachycardia
â– Ventricular: Vent Fib (coarse), Vent Fib (fine), Premature Ventricular
Contraction (PVC) 6/min,
â– PVCs 12/min, PVCs 24/min, Frequent Multifocal, Bigeminy, Trigeminy, Paired
PVCs, Run 5
â– PVCs, Run 11 PVCs, Vent Tachycardia, Asystole
â– Premature Beat: Atrial Premature Atrial Contraction (PAC), Nodal PNC, Left
Ventricular Focus
â– PVC1, Early Left Ventricular Focus PVC1, Right Ventricular Focus PVC1, Early
Right
â– Ventricular Focus PVC1, R on T Left Ventricular Focus, R on T Right
Ventricular Focus, Multifocal PVC
â– Conduction Defects: First Degree Heart Block, Second Degree Heart Block,
Third Degree Heart   Block, Right Bundle Branch Block, Left Bundle Branch
Block
ECG Performance Testing
■ECG amplitude:0.5,1.0,1.5 and 2.0mV
■The accuracy of Amplitude: ±2%
â– Pulse: 30 and 60 BPM, the width of pulse wave is 60 ms
â– Square: 0.125 and 2.0 Hz
â– Triangle: 2.0 and 2.5 Hz
â– Sine: .0.5, 5, 10, 40, 50, 60 and 100 Hz
â– ECG Rate:15, 20, 25, 30, 40, 60, 80, 100, 120, 140, 160, 180, 200, 220, 240,
260, 280, 300,   320, 340 and 350 BPM
■The accuracy of ECG Rate: ±1%
Respiration
â– Respiration Rate: 0 (Apnea), 15, 20, 30, 40, 60, 80, 100, 120 and 150 R/MIIN
■Respiration Impedance: 0.2, 0.5, 1.0 and 3.0 Ω
â– Baseline Impedance:500,1000,1500 and 2000
â– Apnea: 12 s, 22 s, 32 s
â– ECG simulator with adjustable heart rate/12-lead output
â– 33 kinds of Arrhythmia waveforms
â– Pacemaker
â– Two IBP channels
â– Variety of respiratory impedance,apnea pauses (0 BrPM) of 12, 22 and 32
seconds.
â– Standard ECG calibration
â– Export square,triangle and sine with adjustable frequency
IBP
â– The default: 00/00mmHg
â– Dynamic IBP
Channel 1: Artery  120/80mmHg
Radial 120/80mmHg
Left Ventricle 120/00mmHg
Right Ventricle 25/00mmHg
Channel 2: Artery  120/80mmHg
Radial 120/80mmHg
Left Ventricle 120/00mmHg
Right Atrium 15/10mmHg
Right Ventricle 25/00mmHg
Pulmonary Artery 25/10mmHg
Left Atrium 14/4mmHg
Static-Pressure Levels
Channel 1: 0, 80, 160, 240, 320 and 400mmHg
Channel 2: 0, 50, 100, 150, 200 and 250mmHg

Reviews (0)

Reviews

There are no reviews yet.

Be the first to review “Simulator MS400 Patient Multiparameter”

Your email address will not be published. Required fields are marked *